Stratus OCT User Manual PN 2660021134133 A
Specifications
11-3
System Weight
Complete system: 58.6 kg (129 lbs.), comprises Patient Module 24.5 kg (54 lbs.), computer
21.3 kg (47 lbs.), monitor 5.4 kg (12 lbs.), printer 6.4 kg (14 lbs.), keyboard and mouse 0.91
kg (2 lbs.).
Table Loading Capacity
The instrument table is specifically designed for use only with the original Stratus OCT
system components installed in their original configuration. In this original configuration,
the table loading capacity is 63.3 kg (139.5 lbs.) ± 5%.
WARNING: Do not reconfigure system components on the table, nor add
non-system devices or components to the table, nor replace original system
components with substitutes not approved by Carl Zeiss Meditec. Such
actions could result in failure of the table height adjustment mechanism,
instability of the table, tipping and damage to the instrument, and injury to
operator and patient.