Guidance and manufacturer’s declaration - electromagnetic immunity
The Stratus OCT is intended for use in the electromagnetic envir
onment specified below. The customer or user of
the Stratus OCT should assure that it is used in such an environment
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Portable and mobile RF communications
equipme
nt should be used no closer to
any part of the Stratus OCT, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d1.17P=
d1.17P=
80 MHz to 800 MHz
d2.33P=
800 MHz to 2,5 GHz
where P is the maximum output power
r
ating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters,
as determined b
y an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency
range.
b
Interference may occur in the
vicinity of equipment marked with the
following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
Note 1: At 80 MHz and 800 MHz, the higher frequency applies.
Note 2: These guidelines may not apply in all situations. Electr
omagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stat
ions for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM broadcast, cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Stratus OCT is used exceeds the applicable
RF compliance level above, the Stratus OCT should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Stratus
OCT.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Stratus OCT User Manual PN 2660021134133 A
Introduction
1-15