R3132 Series Spectrum Analyzer Operation Manual
5.5 Performance Verification Record Sheet
5-127
For the R3162
For the R3172/82
Measurement a center frequency of 4 GHz
R3162 Setting Specification Result
Attenuator
(dB)
Reference
Level (dBm)
Measured
Value (dBm)
Min. (dB)
Actual
Value (dB)
Max. (dB) Pass/Fail
0 -40 -0.3 +0.3
5 -35 -0.3 +0.3
10 -30 - 0 (Ref) -
15 -25 -0.3 +0.3
20 -20 -0.3 +0.3
25 -15 -0.3 +0.3
30 -10 -0.3 +0.3
35 -5 -0.3 +0.3
40 0 -0.3 +0.3
45 +5 -0.3 +0.3
50 +10 -0.3 +0.3
R3172/82 Setting
Specification
Result
Switching Accuracy Step to Step Accuracy
Attenuator
(dB)
Reference
Level (dBm)
Spec. (dB)
Actual
Value (dB)
Spec. (dB)
Actual
Value (dB)
Pass/Fail
10 -50 0 (Ref) 0 (Ref) 0
20 -40 ±2 ±1.1
30 -30 ±2 ±1.1
40 -20 ±2 ±1.1
50 -10 ±2 ±1.1
60 0 ±2 ±1.1
70 +10 ±2 ±1.1