10 Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
2D-LC User Guide 286
Lab Advisor Service & Diagnostic
Decluster the 2D-LC Cluster
This tool allows to remove an LC device's clustering configuration data, e.g. the
linking between
2
D pump and 2D-LC valve.
1 Select Service & Diagnostic from the menu.
2 Select the
2
D pump.
3 Select Firmware Declustering.
4 To Clear clustering configuration data, press the Run button.
Figure 149 Firmware Declustering of the 2D-LC 1290 binary pump
When
Replacement of one of the cluster partners.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
• Documentation provided with the Agilent Lab Advisor online help
• 2D-LC Manual
Procedure to follow:
• Close the current Acquisition client window
• Close instrument connection from OpenLab Control Panel
NOTE
To re-establish the link between the two modules, re-perform an Auto
configuration and a selection as cluster.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II 2D-LC and is the answer not in the manual?

Agilent Technologies 1290 Infinity II 2D-LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II 2D-LC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals