11 Error Information
Module-Specific Error Messages
2D-LC User Guide 316
External Valve falls into resident mode
Error ID: Flashing status indicator
The valve drive was not able to operate correctly
Probable cause Suggested actions
1 Communication issues
Check the CAN cable connections of the
modules.
Check if the hosted module is present.
2 Configuration mismatch
Check if the firmware on the entire stack is
out of the same firmware set.
Check if the limit of 3 hosted modules for
each host module is not exceeded.
Check if the dipswitch settings are correct.
Check if the firmware on the entire stack has
to be the latest version.