EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>1290 Infinity II 2D-LC

Agilent Technologies 1290 Infinity II 2D-LC User Manual

Agilent Technologies 1290 Infinity II 2D-LC
395 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
2D-LC User Guide 291
10 Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
2D-LC Capillary Leak Test
Leak and tightness check of the 2D-LC Valve with the
2
D pump in the flow path of
the second dimension.
This 2D-LC test works only for the driver-based 2D-LC solution.
1 Select Service & Diagnostic from the menu.
2 Select the
2
D pump.
3 Select the 2D-LC Capillary Leak Test.
4 Press the Run button and follow the instructions in the software.
When
Leak in the 2D-LC valve.
Tools required Description
Wrench, 1/4 - 1/5 inch
Parts required Description
Blank nut
Software required
Agilent Lab Advisor Software (2.18 or higher)
Preparations
Read the following:
• Documentation provided with the Agilent Lab Advisor online help
• 2D-LC Manual
Procedure to follow:
• Close the current Acquisition client window
• Close instrument connection from OpenLab Control Panel

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II 2D-LC and is the answer not in the manual?

Agilent Technologies 1290 Infinity II 2D-LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II 2D-LC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals