2D-LC User Guide 313
11 Error Information
Module-Specific Error Messages
Module-Specific Error Messages
For further module-specific errors, please see the manual of the module in
question.
Initialization of Valve Failed
Error ID: 24000
During the initialization process the motor of the valve drive moves to some
special positions depending on the installed valve head. A failure in this process
means either that the movement couldn’t be performed properly or it was not
noticed correctly by the sensor.
Probable cause Suggested actions
1 Mechanical problems. Friction too high or
blockages on the valve drive’s motor or on
the valve head.
Check valve head for correct installation
Try to identify the source of trouble by
installing a different valve head if possible.
Contact your Agilent Service representative.
2 Defect Sensor on the Valve Drive Motor
Check valve head for correct installation
Try to identify the source of trouble by
installing a different valve head if possible.
Contact your Agilent Service representative.