2D-LC User Guide 313
11 Error Information
Module-Specific Error Messages
Module-Specific Error Messages
For further module-specific errors, please see the manual of the module in
question.
Initialization of Valve Failed
Error ID: 24000
During the initialization process the motor of the valve drive moves to some
special positions depending on the installed valve head. A failure in this process
means either that the movement couldn’t be performed properly or it was not
noticed correctly by the sensor.
Probable cause Suggested actions
1 Mechanical problems. Friction too high or
blockages on the valve drive’s motor or on
the valve head.
• Check valve head for correct installation
• Try to identify the source of trouble by
installing a different valve head if possible.
• Contact your Agilent Service representative.
2 Defect Sensor on the Valve Drive Motor
• Check valve head for correct installation
• Try to identify the source of trouble by
installing a different valve head if possible.
• Contact your Agilent Service representative.