2D-LC User Guide 311
11 Error Information
General Error Messages
Compensation Sensor Open
Error ID: 0081
The ambient-compensation sensor (NTC) on the mainboard in the module has
failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
mainboard is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If the
resistance across the sensor increases above the upper limit, the error message
is generated.
Compensation Sensor Short
Error ID: 0080
The ambient-compensation sensor (NTC) on the mainboard in the module has
failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
mainboard is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If the
resistance across the sensor falls below the lower limit, the error message is
generated.
Probable cause Suggested actions
1 Defective mainboard.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Defective mainboard.
Please contact your Agilent service
representative.