EasyManuals Logo

Agilent Technologies 1290 Infinity II 2D-LC User Manual

Agilent Technologies 1290 Infinity II 2D-LC
395 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #224 background imageLoading...
Page #224 background image
8 Run the System
Run the Experiment
2D-LC User Guide 224
Run the Checkout Familiarization Procedure for ASM
Comprehensive
To run the checkout, various hardware configurations are possible, see Table 5
on page 56. Not all options can be shown. As example Table 14 on page 198 is
used here.
To achieve optimal sensitivity, in comprehensive mode, especially for LC/MS
applications, the LC flow is often split prior to the mass spectrometer. The
following parameters have been optimized for this standard configuration.
Parameters can deviate slightly for your system. Run the experiment with ASM
OFF and subsequently edit or optimize the method for your setup.
Table 31 Example for a MS passive splitter setup (ratio 1:2)
Description (PN) Usage
TEE, ST, 1/16 inch, Low Dead Volume
(0100-0969)
T-piece
DAD Heat Exchanger Capillary 310 mm, 0.12 mm
i.d. (G1315-87339)
2D detector connected to T-piece
Capillary ST 0.075 mm x 500 mm, long socket
(5500-1205)
Inlet of the LCMS source connected to the other
end of the T-piece
Capillary ST 0.075 mm x 250 mm, long socket
(5500-1206)
Remaining connection to the T-piece is used as
waste capillary

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II 2D-LC and is the answer not in the manual?

Agilent Technologies 1290 Infinity II 2D-LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II 2D-LC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals