2D-LC User Guide 287
10 Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
Pump Head Leak Test for the
2
D Pump
The test determines the leakage of the individual pump heads.
This 2D-LC test works only for the driver-based 2D-LC solution.
1 Select Service & Diagnostic from the menu.
2 Select the
2
D pump.
3 Select the Pump Head Leak Test.
Figure 150 Pump Head Leak Test for the 2D-LC 1290 binary pump
4 Press the Run button and follow the instructions in the software.
When
Diagnostic of the
2
D pump.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Procedure to follow:
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel