2D-LC User Guide 289
10 Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
System Pressure Test for the
2
D Pump
The test determines the leak tightness of the system between pump and blank
nut.
This 2D-LC test works only for the driver-based 2D-LC solution.
1 Select Service & Diagnostic from the menu.
2 Select the
2
D pump.
When
Leaks in the system flow path.
Tools required Description
Wrench, 1/4 - 1/5 inch
Parts required Description
Blank nut
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
• Documentation provided with the Agilent Lab Advisor online help
• 2D-LC Manual
Procedure to follow:
• Close the current Acquisition client window
• Close instrument connection from OpenLab Control Panel