10 Troubleshooting and Diagnostics
Lab Advisor Instrument Control
2D-LC User Guide 280
Lab Advisor Instrument Control
2D-LC Hardware License Handling
When
Installation/Deinstallation of USB Hardware Dongle in the
2
D pump of a 2D-LC instrument, to do the
following:
Verify the license status
Verify the correct installation of the USB dongle
De-activate the license on the current module, e.g. to transfer the license to a different pump
module
Parts required p/n Description
5017-0006 USB Dongle
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Procedure to follow:
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
NOTE
The
2
D pump must be a 1290 Infinity I, II or 1290 Infinity II biocompatible Binary
Pump.