EasyManua.ls Logo

Agilent Technologies 1290 Infinity II 2D-LC

Agilent Technologies 1290 Infinity II 2D-LC
395 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Troubleshooting and Diagnostics
Lab Advisor Instrument Control
2D-LC User Guide 284
Instrument Control of the 2D-LC Cluster
1 Select Instrument Control of the
2
D pump (2D-LC cluster).
2 Change the settings of the
2
D pump as required.
3 To identify a valve, select the valve from the Valve drop-down list.
The following instrument setups are possible:
One 2D-LC valve
Three valves:
One 2D-LC valve
Two MHC valves
Figure 146 Example of a 2D-LC instrument with a 2D-LC ASM valve and two MHC valves
When
Control the behavior of the
2
D pump and the 2D-LC valves.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Procedure to follow:
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
NOTE
To use instrument control of the
2
D pump, the 2D-LC hardware license must be
active.

Table of Contents

Related product manuals