2D-LC User Guide 315
11 Error Information
Module-Specific Error Messages
Pressure Cluster Partner Missing
The connection from the valve drive to a defined pressure cluster partner is lost.
Position Cluster Partner Missing
Probable cause Suggested actions
1 Communication issues
Check the CAN cable connections of the
modules.
2 Configuration mismatch
Check and correct if necessary the valve
configuration and presence of defined pressure
cluster partner.
Probable cause Suggested actions
1 Communication issues
Check the CAN cable connections of the
modules.
2 Configuration mismatch
Check and correct if necessary the valve
configuration and presence of defined position
cluster partner.