EasyManuals Logo

Agilent Technologies 1290 Infinity II 2D-LC User Manual

Agilent Technologies 1290 Infinity II 2D-LC
395 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
2D-LC User Guide 315
11 Error Information
Module-Specific Error Messages
Pressure Cluster Partner Missing
The connection from the valve drive to a defined pressure cluster partner is lost.
Position Cluster Partner Missing
Probable cause Suggested actions
1 Communication issues
Check the CAN cable connections of the
modules.
2 Configuration mismatch
Check and correct if necessary the valve
configuration and presence of defined pressure
cluster partner.
Probable cause Suggested actions
1 Communication issues
Check the CAN cable connections of the
modules.
2 Configuration mismatch
Check and correct if necessary the valve
configuration and presence of defined position
cluster partner.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II 2D-LC and is the answer not in the manual?

Agilent Technologies 1290 Infinity II 2D-LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II 2D-LC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals