2D-LC User Guide 119
5 2D-LC Data Acquisition in MassHunter Workstation 11
Tune Window
Tune Window
In the Tune window, you tune the mass spectrometer. You can use one of the
automated tuning algorithms, or you can manually tune the instrument. Manual
tuning can result in a less than optimal tune; however, if you perform a manual
tune, Agilent recommends that you only manually tune the front part of the
instrument: ion source and optics 1. Agilent does not recommend that you
manually tune parameters that are after the collision cell.