Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 115
4 Removal/ Installation, Replacement and Adjustment
Pre Optics System
Axial window
1 Remove the 3 skimmer cone nuts and remove skimmer cone.
2 Withdraw axial window.
3 Unscrew window collet from assembly and replace or clean window as required.