Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 73
4 Removal/ Installation, Replacement and Adjustment
Front left lower cover
Front left lower cover
1 Loosen screw in sample compartment.
2 Loosen screw at left front of instrument.
3 Slide cover out and up to release tabs at bottom of cover.
Torch compartment door
1 To remove door slide up and out.