Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 81
4 Removal/ Installation, Replacement and Adjustment
Peristaltic Pump
6 Replace and reconnect hoses. Check for leaks before securing interface.
Peristaltic Pump
1 The peristaltic pump is part of the sample introduction system providing sample uptake flow
and waste flow. There are both 3 and 5 channel pumps available depending on the
configuration. Both pumps assemble/disassemble into the instrument in the same manner.
2 To remove the pump assembly remove 4 phillips head screws and withdraw the module
from the chassis.
3 Disconnect cables and replace pump as required.