Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 15
1 Safety Information
Handling of Gas
Handling of Gas
Warning – Storing Argon Safely
Argon, which is used to create the plasma, is a dangerous gas only if it displaces the air you
are breathing or is mishandled in its storage cylinder. Take the following precautions to
prevent an explosion or suffocation hazard:
• Secure the cylinder valve caps and move the cylinder with an approved handcart.
• Label the gas cylinder clearly and store it vertically away from all ignition and heat sources.
• The Argon cylinder will release argon through a pressure-relief device if the temperature
exceeds 52°C.
• Protect cylinders that are stored outside from the sun and extreme temperatures. Place
the cylinder on a level surface above ground.
• Attach the argon hoses tightly to both the instrument and gas source. Route the hoses so
they cannot be damaged or crimped. Check for leaks using an electronic leak detector or
by pressurized test.
• Provide adequate ventilation around the gas cylinders, especially if they are placed in a
small storeroom.
Warning – Other Gases
Gases must be handled with care.
• Refer to the MSDS (Material Safety Data Sheet) for safe handling information.
• Check for gas leaks periodically using the appropriate tools (leak detector, pressure test,
etc.)