82 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Advanced Valve System (AVS)
Advanced Valve System (AVS)
The Advanced Valve System is an optional accessory available to increase speed of sample
throughput and reduce sample inject volume.
There are 3 types available, 4 port, 6 port and 7 port. All versions remove and install in the
following manner.
1 Pull tray forward to remove from locating pins.
2 Remove 4 Phillips head screws (2 point). Support assembly when removing 4
th
screw.
3 Slide assembly forward being careful not to catch the Argon bubble inject hose on the
instrument chassis.