Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 135
4 Removal/ Installation, Replacement and Adjustment
Polychromator
15 Remove the screws retaining the camera connection board to the casting.
16 Remove the 2 screws that secure the D range connector to the board.
17 Using a 7 mm nut driver, remove the three Nylock nuts that hold the camera assembly in the
casting.
18 Carefully remove the camera assembly.