Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 173
6 Customer Diagnostics and Problem Solving
Testing Nebulizer Back Pressure
Nebulizer Back Pressure failures can be caused by the following:
High Back Pressure: Nebulizer or Torch injector tube blocked.
Low Back Pressure: Nebulizer leaking or not connected.