Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 25
3 General Information
Environmental Conditions and Utility Requirements
Water cooling maximum inlet pressure (kPa) 400 kPa (58 psi)
Water cooling fittings for pipe (Chiller end) 12mm barbs with hose clamps
Water cooling fittings for pipe (Instrument end) 12mm compression fitting
Temperature and Humidity ranges
Temperature drift over 1-hour period 15 to 30°C <2°C change in 1 hour
Relative humidity 20 to 80% (non condensing)
Altitude Up to 3000m
Atmosphere Non condensing, Non corrosive
Storage Conditions
Temperature 5 to 60°C
Relative humidity 15 to 85% (non condensing)
Agilent 5110 ICP-OES Specification