Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 305
12 Peripherals
Routine maintenance
Preventative Maintenance visit:
1 Disconnect the return hose at the water cooler.
2 Turn the water cooler on; empty the water into a suitable container and dispose of it.
3 Turn the water cooler off and disconnect from mains supply. Locate and remove the strainer.
Refer to the water cooler operator instruction booklet for further details on the location of the
strainer.
4 Rinse the strainer in clean water and refit to the water cooler. Connect the return supply hose
to the water cooler and fill with cooling water.