164 Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
9 Faults and Errors
Front (or Back) turret error Something has interfered with the
turret rotation.
The turret motor/encoder assembly
is inoperable.
The turret type was changed while
the power was on and the turret
alignment procedure was not
performed.
Turret is loose.
1 Clear any obstructions.
2 Check the Align Mode light. If it is
on, perform the alignment
procedure. See Align Mode on
page 150.
3 Tighten the knurled nut on the top
of the turret.
4 If the error occurs again, obtain
Agilent service.
Injector not present There is a board failure in the
injector or GC.
The injector cable is bad or not
securely connected to the GC.
There is a cable failure in the GC.
Your method specifies an incorrect
injector location (method
mismatch).
1 Make sure that the injector-to-GC
cable connection is secure.
2 Check your method to make sure it
uses the appropriate injector
location.
3 If the error remains, obtain Agilent
service.
Injector offline There is a board failure in the
injector or GC.
The injector cable is bad or not
connected.
There is a cable failure in the GC.
1 Make sure that the injector to GC
cable connection is secure.
2 If the error remains, obtain Agilent
service.
No Bar Code Reader Bar Code Reader cable not securely
connected.
Bar Code Reader is defective.
Tray is defective.
1 Ensure the Bar Code Reader cable
connection is secure.
2 If the problem continues, obtain
Agilent service.
No bottle in gripper The sample vial was not found by
the gripper.
The gripper could not grasp the vial.
The vial was dropped during
transfer to or from the turret.
The sensor in the gripper is
defective.
The vial does not meet
specifications.
1 Make sure that the sample vials are
in the locations specified by the
sequence.
2 Make sure that the sample vials
meet recommended specification.
3 If you are using adhesive labels,
verify that the labels are properly
installed. See “Labeling sample
vials” on page 112.
4 If the error recurs frequently, obtain
Agilent service.
Table 2 5 Error messages (continued)
Message Probable cause Suggested action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies agilent 7683B and is the answer not in the manual?

Agilent Technologies agilent 7683B Specifications

General IconGeneral
BrandAgilent Technologies
Modelagilent 7683B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals