98 Installation, Operation, and Maintenance
Part 2, Operation
4 Controlling the Sampler
Remember that the solvents you use (A and/or B) are set within the method.
See “Wash parameters" on page 95.
Waste bottle usage
For 3-sample turrets only. If the waste bottle control does not appear on the
screen, press More to select that option. The positions containing bottles to
use for waste (A, B, or Alternate between A and B).
Solvent volume
Solvent volume allows you to conserve solvent by adjusting the amount used to
wash the syringe.
Enable or disable saver mode and specify the amount of solvent to draw into
the syringe on each wash. The default, if disabled, is 80% of the syringe
volume.
Injector fan
If the fan control does not appear on the screen, press More to select that
option. Normally, leave the fan on to help cool the injector and keep your
samples stable. Note that the fan will briefly turn off once per day for a short
time. This extends fan life.