G4208A User’s Guide 110
3 Running an Isocratic Analysis
Preparing the LC System
Preparing the LC System
1 For the isocratic pump, fill the solvent bottle with the mixture of LC- grade
bidistilled water (35
%) and acetonitrile (65 %). For the binary or quaternary
pump, fill one solvent bottle with bidistilled water (channel A) and the other
with acetonitrile (channel B).
2 Turn on the detector lamp and pump via the Control - System On/OFF screen.
3 For the quaternary pump, turn on the degasser by pressing the line-power
switch.
4 Purge the pump.
5 Allow the detector at least 15 min to provide a stable baseline.
6 Transfer the contents of an Agilent Technologies isocratic standard sample
ampoule into a vial and seal the vial with a cap. Place the vial in position 1 of
the autosampler tray.
7 Pump the water/acetonitrile (35/65 %) mobile phase through the column for
10
min at a flow rate of 2 mL/min.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals