G4208A User’s Guide 120
4 Running Multiple-Vial Analyses
Single-Level Calibration Sequences
Single-Level Calibration Sequences
The following procedure describes how to set up a calibration sequence for an
analysis which uses single- level calibration.
There is one calibration standard (C) and 9 samples (S).
The analysis requires that:
• each sample is analyzed in duplicate,
• the calibration standard is analyzed once before the samples and re- analyzed
once after every 2 samples,
C SS C SS C SS C SS C S C
• the calibration standard is located in position 90 and the 9 sample vials are in
positions 1 to 9 of the autosampler tray, and
• the method called METH1 is used for the samples and standards.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the Delete button to delete all lines or use File - New.
1 Select the Sequence screen.
2 Move to line 1 and press Insert.
3 Select Method and then move to METH1 and press Load.
4 Move to the End of sequence list, select the Wizard and enter:
5 Select Calibration to display the Calibration Settings screen.
Vial Range 1 – 9
#Inj. 2