G4208A User’s Guide 46
1 Start-up Information
Maintenance Information
Maintenance Information
1 To view/change the Maintenance information, press More from the Welcome
screen and select Maintenance from the menu.
Figure 26 Maintenance Screen
The Maintenance System screen shows a list of all modules in the system with
their names, product and serial numbers, and the firmware revision.
2 You can update the firmware using Update Wizard, which allows updating all
modules of the system at once, or using Single to update a selected module.
The firmware must be on an inserted and activated USB Flash Drive in its root
directory.
On the module- specific screens you can
• see the Early Maintenance Feedback (EMF), error and maintenance
events,
• set the EMF limits (see “Early Maintenance Feedback (EMF)” on page 48),
• do module maintenance (e.g. calibrations),
• add maintenance activities into the permanent log,
• identify the module in the stack (flashing LED).
The Instant Pilot provides basic maintenance and diagnostic functions only. The
Agilent LabAdvisor software provides the full maintenance and diagnostic
capabilities.
update a single module
update a set of modules
to change the product number
or serial number after main
board exchange
leave this screen
system or module specific
information