G4208A User’s Guide 134
4 Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices
NOTE
An Agilent variable wavelength detector or diode array detector will perform a
balance (assumed that Auto Balance is set to pre-run) when receiving a start
command from the Instant Pilot. This will only happen in the Standard and Wait
for single (multiple) start request modes. In the Send single (multiple) start
request modes, a balance before the run will NOT be performed. If regular
balancing is required, set the Auto Balance check box to Postrun.