Configuring the Heartstream XL
10-8 Setting Up and Configuring the Heartstream XL
To turn the Heartstream XL on in Configuration Mode:
1. If the device is already on, turn the Energy Select knob to Off.
2. While holding down softkeys 4 & 5, turn the Energy Select knob
to AED On.
The configuration menu appears as shown in Figure 10-6. The menu lists the
categories of settings that may be configured.
)LJXUH&RQILJXUDWLRQ0HQX
Configurable Parameters
The following tables show the configurable parameters for each category of
settings. A description of each parameter is provided, along with the possible
choices. Default settings are in bold.
ENTER
MAIN
AED Settings
Manual Settings
ECG Filter Settings
Save Settings to Data Card
Load Settings from Data Card
Print All Settings
Turn Off To Exit Configuration
General Settings