Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-5
13
ECG Monitoring
Inputs: Single channel ECG may be viewed on display and printed. Pads
ECG is obtained through 2 multifunction defibrillation electrode pads. Lead
I, II, or III is obtained through the 3 lead ECG cable and separate monitoring
electrodes. With a 5 lead cable, lead aVR, aVL, aVF, or V can also be
obtained.
Lead Fault: LEADS OFF message and dashed line appear on the display if
an electrode or lead wire becomes disconnected.
Paddle Fault: NO PADDLES CONNECTED message and dashed line
appear on the display if paddles become disconnected.
Pad Fault: PADS OFF message and dashed line appear on the display if a pad
becomes disconnected.
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an
accuracy of ±10%.
Heart Rate Alarms: Configurable pairs of low and high heart rate alarm lim-
its: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Hands Free Defibrillation Patient Cable Length: 7 ft (2.13 m).
ECG Cable Length: 12 ft (3.7 m).
Common Mode Rejection: Greater than 90 dB measured per AAMI standard
for cardiac monitors (EC 13).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals