Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-9
13
Thermal Array Printer
Continuous Real Time Strip: User starts and stops the strip. The strip prints
the selected ECG lead with the following data:
Header 1: Date, Time, Heart Rate, the SPO2 Value (if available), and the text
"Delayed" if printing has been configured for Delayed Mode. Prints every 12
seconds.
Header 2: Current mode (AED/Manual), Lead, Gain, filter setting, the text
"Sync" (if Sync has been enabled), and Pacer Settings (consisting of the Pacer
Mode, Rate, and Current, if presently pacing the patient). Prints every 12 sec-
onds, with Header 1.
Header 3: Changes in Mode, Gain, Lead, Sync, and Pacer Settings.
Footer: Drug Annotations, HR/SpO2 limits on a Limit Alarm, the Results of
Analysis in AED Mode (No Shock Advised, Shock Advised, or Cannot Ana-
lyze), Charging to xxx J, Shock Delivered, No Shock Delivered, Disarm, Bat-
tery Low.
Symbols: Mark Triangle (for presses of the Mark key), an Alarm Bell (Alarm
Limit Violations), Lightning Bolt (Shock Delivered; followed by b for bipha-
sic), Vertical Stripe Boundaries/Pacer/Sync Tick Marks.
Event Printing: Mark key automatically documents ECG and events during
defibrillation episodes. The Mark key can annotate the event with one of the
following labels: Epinephrine (Adrenaline), Atropine, Lidocaine, and Other.