System Messages
12-2 Troubleshooting
System Messages
System messages remain on the display until the specified action is taken or
no longer relevant. Each new message displayed is accompanied by three
beeps to alert you. Table 12-1 lists system messages.
7DEOH6\VWHP0HVVDJHV
Message Description Corrective Action
&RQIL JXUDWLRQ /RVW The configuration is reset to the default
settings.
l
Reconfigure the Heartstream XL.
l
Check to see if the battery is properly
charged.
l
Replace the battery.
l
If the message persists, call for
service.
'DWD&DUG'LVDEOHG The Data Card is not in use because it is
full, incompatible, absent, or inserted
after the Heartstream XL was turned on.
If possible, turn the Heartstream XL off
for more than 2 minutes, insert an empty,
compatible Data Card, and turn the
device on.
You may also enter configuration mode
and turn the machine off, then on again.
(&*)DXOW The ECG data acquisition system failed
and data is unavailable from the 3- or 5-
wire monitoring electrodes.
Remove the device from active use and
call for service.
/RZ%DWWHU\ The battery has sufficient capacity
remaining to provide only about ten min-
utes of monitoring time and 5 shocks
before the Heartstream XL shuts off.
l
Replace the battery with a fully
charged battery.
l
Plug in AC power.