Defibrillating in Manual Mode
M4735A Heartstream XL Defibrillator/Monitor 6-3
6
Defibrillating in Manual Mode
This section will explain how to prepare for and perform asynchronous
defibrillation in Manual Mode using multifunction defib electrode pads, and
paddles.
127( In Manual Mode, defibrillation is always performed through paddles or pads.
However, during defibrillation, you may choose to monitor leads acquired
through an alternate ECG source (3- or 5-lead monitoring electrodes).
Using External Paddles
In preparation for defibrillation in Manual Mode using external paddles, per-
form the following steps:
1. If needed, insert a Data Card (as described in “Using a Data Card
(Optional)” on 2-13).
2. Turn the Energy Select knob to Manual On.
3. Remove paddles by simultaneously pulling them up and out of
the holders.