13-8 Specifications & Safety
Display
Size: 115 mm x 86 mm
Type: EL - Electroluminescent
Resolution: 320 x 240 pixels
Sweep Speed: 29 mm/s nominal (stationary trace;
sweeping erase bar).
Viewing Time: 4 seconds.
Battery
Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid.
Dimensions: 2.4" (H) x 0.94" (W) x 7.2" (D) (61.7mm x 23.9mm x 182mm).
Weight: 1.4 lb. (0.65 kg).
Charge Time: Approximately 14.5 hours to 100%. Approximately 3 hours to
90%, indicated by LED on front panel.
Capacity: 100 minutes ECG monitoring or 50 full-energy discharges or 75
minutes ECG monitoring while pacing (with a new, fully charged battery at
room temperature, 25º C).
Battery Indicators: LOW BATTERY message appears on display when at
least 10 minutes of monitoring time and 5 maximum energy discharges
remain (with a new battery at room temperature, 25º C).
Battery Storage: Should not be stored above 40º C for extended periods of
time.
Charger Output: Unit can be operated using only AC Power, with no battery
installed.