EasyManuals Logo

Agilent Technologies M4735A User Manual

Agilent Technologies M4735A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-7
13
(&*$QDO\VLV3HUIRUPDQFH
5+<7+0&/$66
(&*7(67
6$03/(
$
6,=(
120,1$/63(&,),&$7,216
Shockable Rhythm —
ventricular fibrillation
600 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(sensitivity >90%) for adult defibrillation
Shockable Rhythm —
ventricular tachycardia
300 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(sensitivity >75%) for adult defibrillation
Non-Shockable Rhythm —
Normal Sinus Rhythm
250 Meets AAMI DF39 requirement (specificity > 95%) and
AHA recommendation
b
(specificity >99%) for adult
defibrillation
Non-Shockable Rhythm —
Asystole
500 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(specificity >95%) for adult defibrillation
Non-Shockable Rhythm —
All other non-shockable
rhythms
600 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(specificity >95%) for adult defibrillation
a. From Agilent Technologies ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy.
Automatic External Defibrillators for Public Access Use: Recommendations for Specifying and
Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New Waveforms, and
Enhancing Safety. American Heart Association (AHA) AED Task Force, Subcommittee on AED
Safety & Efficacy. Circulation 1997;95:1677-1682.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals