Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-7
13
(&*$QDO\VLV3HUIRUPDQFH
5+<7+0&/$66
(&*7(67
6$03/(
$
6,=(
120,1$/63(&,),&$7,216
Shockable Rhythm —
ventricular fibrillation
600 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(sensitivity >90%) for adult defibrillation
Shockable Rhythm —
ventricular tachycardia
300 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(sensitivity >75%) for adult defibrillation
Non-Shockable Rhythm —
Normal Sinus Rhythm
250 Meets AAMI DF39 requirement (specificity > 95%) and
AHA recommendation
b
(specificity >99%) for adult
defibrillation
Non-Shockable Rhythm —
Asystole
500 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(specificity >95%) for adult defibrillation
Non-Shockable Rhythm —
All other non-shockable
rhythms
600 Meets AAMI DF39 requirement and AHA recommenda-
tion
b
(specificity >95%) for adult defibrillation
a. From Agilent Technologies ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy.
Automatic External Defibrillators for Public Access Use: Recommendations for Specifying and
Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New Waveforms, and
Enhancing Safety. American Heart Association (AHA) AED Task Force, Subcommittee on AED
Safety & Efficacy. Circulation 1997;95:1677-1682.