Specifications
13-12 Specifications & Safety
Data Card Event Summary: The Data Card stores continuous ECG wave-
forms and events on a Type II PCMCIA card, SanDisk SDP3B 8MB ATA
FlashDisk. One 8MB card can store approximately 2 hours of continuous
ECG waveforms and events.
General
Dimensions: 19.0 cm(H) x 37.6 cm(W) x 34.6 cm(L)
(7.5 in. x 14.8 in. x 13.7 in.)
Weight: Standard Configuration weighs 6.5 kg (14.3 lbs.) including battery,
full roll of paper, and external paddles.
Environmental
Temperature: 0º C to 55ºC operating, -20º to 70ºC storage.
l
Charging the battery at temperatures above 35ºC may degrade bat-
tery life
l
Storing the battery for extended periods at temperatures above 40ºC
will reduce battery capacity and degrade battery life.
Humidity: Up to 95% Relative Humidity
l
Printer paper may jam if paper is wet.
l
Thermal Printer may be damaged if wet paper
is allowed to dry while in contact with printer
elements.