Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-13
13
Altitude:
Operating: up to 15,000 ft.
Storage: up to 15,000 ft.
Shock: Agilent Technologies, Section 760 Class B1 Drop Test
(200 G’s, < 3 msec pulse).
Vibration: Agilent Technologies, Section 759 Class B1 Vibration.
Water Resistance: Meets IEC 601-2-4, IPX0.
EMC: Meets EN 60601-1-2.
Safety: Meets IEC 601-1 (EN 60601-1), UL 2601-1, CAN/CSA C22.2 No.
601-1.
Other Considerations: Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.
AC Line Powered: 100-240 VAC, 50/60 Hz, 1.5 A (Class 1)
Battery Powered: 12 V Rechargeable, SLA