Defibrillation Therapy
M4735A Heartstream XL Defibrillator/Monitor 1-5
1
Indications for AED Therapy
An AED is to be used in the presence of a suspected cardiac arrest on patients
of at least 8 years of age that are:
l
Unresponsive
l
Not breathing
l
Pulseless
Contraindications for AED Therapy
An AED is not to be used on patients that exhibit one or any combination of
the following:
l
Responsiveness
l
Spontaneous breathing
l
Palpable pulse
Precautions for AED Therapy
The AED algorithm is not designed to handle erratic spiking problems caused
by a properly or improperly functioning pacemaker. In patients with cardiac
pacemakers, the M4735A Heartstream XL Defibrillator/Monitor may have
reduced sensitivity and not detect all shockable rhythms.
NOTE The AED algorithm is not intended for children under 8 years of age. For chil-
dren older than 8 years, the American Heart Association recommends that
standard operating procedures for AEDs be followed. American Heart Asso-
ciation Textbook of Advanced Cardiac Life Support. Dallas, Texas.:
AHA; 1997.