EasyManuals Logo

Agilent Technologies M4735A User Manual

Agilent Technologies M4735A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
Changing Pacing Modes
8-8 Pacing (Optional)
Changing Pacing Modes
If paced pulses are being delivered, you must stop pacing before changing the
pacing mode. For example:
1. Press to stop pacing.
2. Press to change the Mode.
3. Adjust the rate, if needed.
4. Press to resume pacing.
5. Adjust the output, as needed to obtain capture.
Defibrillating During Pacing
If the patient must be defibrillated during pacing, follow the procedure for
defibrillating in Manual Mode on page 6-1.
Pacing is automatically turned off when you charge the defibrillator and the
pacing dialogue box is removed from the display. After a shock, pacing
remains off.
To resume pacing, refer to “Performing Pacing” on page 8-5. When you
resume, settings selected prior to defibrillation (rate, mode, and output)
remain the same.
Start
Stop
Mode
Start
Stop

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals