Performing Pacing
M4735A Heartstream XL Defibrillator/Monitor 8-7
8
6. Increase the output until cardiac capture occurs.
Press on to increase the output in increments of 10 mA.
7. Decrease the output to the lowest level that still maintains capture.
Press on to decrease the output in increments of 5 mA.
8. Press to stop pacing.
9. Press to exit the pacing function. The green LED next to the but-
ton goes out, indicating pacing is no longer active.
127( If pacing using pads, pacing will not start if there is a problem with the multi-
function defib electrode pads connections. If in demand mode, pacing will not
start if there is a problem with the ECG monitoring electrodes connections. In
either case, if a problem occurs a system message is displayed.
127( The pacing window remains on as long as the pacing function is enabled.
Output
Output