Performing Pacing
8-6 Pacing (Optional)
3. Press to change to fixed mode, if in demand mode or if R-wave
detection is unreliable.
The pacing dialogue box displays the current mode. To switch back to
demand mode, press again.
4. Adjust the rate to the desired number of paced pulses per minute (ppm).
Press up ( ) or down ( ), on to increase or decrease the
number of paced pulses per minute.
5. To start pacing, press .
The message 3DFLQJ indicates that paced pulses are being delivered in the
selected mode at the rate and output level displayed.
)LJXUH3DFLQJZLWK3DGV'LVSOD\
Mode
Rate
Start
Stop
HR ALARM
LEAD
SELECT
CHARGE
Shocks:0
00:00:8
78
Pads
Pacer Stop
Demand Mode
70ppm 30mA
SpO2
ON/OFF
SYNC
ON/OFF