Specifications
13-2 Specifications & Safety
Specifications
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance.
Shock Delivery: Via multi-function defib electrode pads or paddles.
Delivered Energy Accuracy:
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged
M3516A SLA battery pack at 25º C. Less than 15 seconds to 200 Joules,
when powered by AC with no battery installed.
Patient Impedance Range:
Minimum: 10-25 Ohm, depending upon energy level
Maximum: 180 Ohm