Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-3
13
Manual Mode
Manual Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100,
150, 200 Joules.
Controls: Manual/AED On/Energy Select knob, Charge/Disarm, Shock,
ECG Lead Select, SpO2 On/Off, SpO2 Alarm, HR Alarm, Sync On/Off,
Pacer, Pacer Start/Stop, Pacer Rate, Pacer Current, Pacer Mode, ECG
Gain, Volume, Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
QRS Beeper, Charging tones (for sync and asychronous modes), AC
Power LED, Battery Charging LED, Sync LED, Pacer LED.
Armed Indicators: Charge done tone and available energy indicated on
display.
Energy Selection: Front panel rotary knob.
Charge Control: Front panel "2" key or buttons on paddles.
Shock Control: Front panel "3" key or buttons on paddles.
Synchronizer: SYNC message appears on the monitor and is annotated on
the printer (if printing while in Sync mode) An audible beep sounds with
each detected R-wave, while a tick mark on the monitor and printed strip
indicate the discharge points. Synchronizer delay is less than 60 msec from
peak R-wave to peak current of the defibrillation discharge.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals