Specifications
13-4 Specifications & Safety
AED Mode
AED Energy Profile: Fixed energy (150 Joules).
AED Shock Series: 2, 3, or 4 shocks per series.
Shock Series Timer: off, 30, 60, 90, 120, 150, 180,
or 210 seconds.
Text and Voice Prompts: Extensive text/audible
messages guide user through protocol.
AED Controls: On, Off, Pause/Resume, Analyze/Stop Analysis, Shock,
Lead Select, SpO2 On/Off, SpO2 Alarms, HR Alarms, ECG Gain, Vol-
ume, Strip, Summary, Mark.
Indicators: EL display for ECG waveform and text prompts, Audio alerts,
Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone, Printer,
AC Power LED, Battery Charging LED.
Armed Indicators: Charge Done Tone, Available Energy indicated on
display, Voice Message.
Patient Analysis: Per protocol, evaluates patient ECG and signal quality
to determine if a shock is
appropriate and evaluates connection impedance for proper defibrillation
pad contact.
Shockable Rhythms: Ventricular Fibrillation with amplitude>100 uV and
wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity: Meets AAMI guidelines.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals