Printing Events
M4735A Heartstream XL Defibrillator/Monitor 9-9
9
Printing Events
The Heartstream XL can be configured to print automatically when certain
events occur. The table below lists these events and the length of the strip
printed, depending on whether the printer is configured to print real-time or
with a 6-second delay.
7DEOH&RQILJXUDWLRQVIRU/HQJWKRI3ULQWHG6WULSV
Event
Real-Time
Strip Length
Delayed
Strip Length
Defibrillator charges continuous 6 seconds just prior to charging, plus
continuous printing throughout the
charge duration.
Shock Delivered 12 seconds 6 seconds just prior to shock, plus 12
seconds after shock.
Shock Failed 6 seconds 6 seconds just prior to the message
1R6KRFN'HOLYHUHG, plus 6 seconds after
the message.
Defibrillator disarmed 6 seconds 6 seconds just prior to disarm, plus
6 seconds after disarm.
SpO2 or Heart Rate
Alarm Violation
6 seconds 6 seconds just prior to alarm viola-
tion, plus 6 seconds after alarm
violation.
pressed
6 seconds 6 seconds just prior to marking, plus
6 seconds after button pressed.
Mark