Specifications
M4735A Heartstream XL Defibrillator/Monitor 13-11
13
SpO2 Pulse Oximetry
Accuracy with:
M1191A sensor - 1 standard deviation 70% to 100%, ± 2.5%.
M1192A sensor - 1 standard deviation 70% to 100%, ± 2.5%.
M1194A sensor - 1 standard deviation 70% to 100%, ± 4.0%.
NELLCOR sensors - 1 standard deviation 80% to 100%, ± 3.0%.
Resolution: 1%.
Sp02 Alarm Limits: Three preset low alarm limits: 90%, 85%, and 80%.
INOP Alerts: Triggered by disconnected sensor, noisy signal, light interfer-
ence or low signal (non-pulsatile).
Event Storage
Internal Event Summary: The internal Event Summary stores up to 300
events and up to 50 waveforms.
Events can be marked with a Mark symbol and, if configured for drug annota-
tion, the following labels can be added: Epinephrine (Adrenaline in U.K. &
Australia), Atropine, Lidocaine, or Other.
The Summary key on the front panel is used to print the internal Event Sum-
mary.