Defibrillating in Manual Mode
M4735A Heartstream XL Defibrillator/Monitor 6-7
6
2. Charge
Press or charge button on paddles.
As the defibrillator charges, the current charge is displayed above the
shock counter (if configured on) as shown in Figure 6-5. A charging tone
beeps until the desired energy level is reached, at which point you’ll hear a
continuous charge tone.
)LJXUH&KDUJLQJ'LVSOD\
You may increase or decrease the selected energy level after pressing the
button. Simply move the Energy Select knob to desired energy
level as before.
The defibrillator charges to the selected energy level automatically.
:$51,1* :DLWXQWLOWKHFX UUHQWFKDUJHUHDFKHVWKHVHOHFWHGHQHUJ\OHYHOEHIRUHUHDGMXVWLQJ
WKHVHOHFWHGHQHUJ\OHYHO
&+$5*(
Pulse
HR ALARM
LEAD
SELECT
SpO2
ON/OFF
6KRFNV
SYNC
ON/OFF
Pads
DISARM
Sp02
ALARM
47J
&+$5*(