System Messages
12-6 Troubleshooting
'DWD&DUG)XOO
l
The incident is more than 2 hours in
duration, causing the Data Card to fill.
l
An empty Data Card was not inserted
for the incident, causing the Data Card
to fill sooner.
l
None. A new Data Card can not be
inserted during an incident.
l
Use one Data Card per incident/
patient to decrease the chance of the
card filling.
'DWD&DUG,QWHUUXSWHG The Data Card is removed during an inci-
dent.
l
None. The Data Card can not be re-
inserted during an incident.
l
Do not remove the Data Card during
an incident.
'DWD&DUG1RW,Q6HUYLFH The Data Card is inserted while the Heart-
stream XL is on.
None. A Data Card must be inserted prior
to turning the Heartstream XL on for the
current patient.
,QFRPSDWLEOH'DWD&DUG A Data Card other than the M3510A is
inserted.
Use only M3510A Data Cards.
1R'DWD&DUG3UHVHQW A Data Card is not in the Heartstream XL. Turn the Heartstream XL off and insert a
Data Card prior to the first event for the
patient.
.H\,QDFWLYH
The key pressed is currently inactive (i.e.
is inactive in AED Mode).
Use the appropriate mode for the key.
6WRS3DFHU
is pressed while pacing pulses
are being delivered.
Stop pacing before changing the pacing
mode.
$WWDFK&DEOH Patient cable is not properly attached to
Heartstream XL.
Check cable connections.
7DEOH&RQWLQXHG0RPHQWDU\0HVVDJHV
Message Possible Cause Corrective Action
Pacer
Mode